Title : 
Traceability of capacitance measurements at NMI VSL
         
        
        
            Author_Institution : 
NMi Van Swinden Lab., Delft
         
        
        
        
        
        
            Abstract : 
NMi VSL is working on its independent realization of the unit of capacitance (F). The traceability will be derived from the DC quantum hall resistance standard. This paper focuses on a quadrature bridge that has recently been completed. The first results obtained from the bridge show an agreement with the NMi VSL reference group of capacitors within 1 muF/F.
         
        
            Keywords : 
bridge instruments; capacitance measurement; electric resistance measurement; measurement standards; quantum Hall effect; units (measurement); DC quantum hall resistance standard; NMi VSL reference group; capacitance measurement; quadrature bridge; traceability chart; unit-of-capacitance; Bridge circuits; Capacitance measurement; Capacitors; Electrical resistance measurement; Impedance; Laboratories; Manufacturing; Oil insulation; Quantum capacitance; Resistors;
         
        
        
        
            Conference_Titel : 
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
         
        
            Conference_Location : 
Broomfield, CO
         
        
            Print_ISBN : 
978-1-4244-2399-6
         
        
            Electronic_ISBN : 
978-1-4244-2400-9
         
        
        
            DOI : 
10.1109/CPEM.2008.4574967