Title :
Calibration of a 3.5 mm coaxial power sensor using a transfer standard with 7 mm coaxial or WR42 waveguide connectors
Author :
Kang, Tae-Weon ; Kim, Jeong-Hwan ; Kwon, Jae-Yong ; Park, Jeong-Il ; Lee, Joo-Gwang
Author_Institution :
Div. of Phys. Metrol., Korea Res. Inst. of Stand. & Sci., Daejeon
Abstract :
A calibration system of a 3.5 mm coaxial power sensor was implemented using transfer standards with 7 mm coaxial or K-band waveguide input connectors. A 7 mm-3.5 mm or WR42-3.5 mm adapter was used to form a combined transfer standard. The effective efficiency of the combined transfer standard was obtained. The 3.5 mm coaxial power sensor was calibrated using the well-known direct comparison method and measurement uncertainty was evaluated.
Keywords :
calibration; electric sensing devices; measurement uncertainty; power measurement; transfer standards; waveguide couplers; K-band waveguide input connector; WR42 adapter; WR42 waveguide connector; coaxial power sensor calibration; measurement uncertainty; size 3.5 mm; size 7 mm; transfer standard; Calibration; Coaxial components; Connectors; Electromagnetic waveguides; K-band; Measurement uncertainty; Metrology; Power measurement; Sensor systems; Thermistors;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 2008. CPEM 2008. Conference on
Conference_Location :
Broomfield, CO
Print_ISBN :
978-1-4244-2399-6
Electronic_ISBN :
978-1-4244-2400-9
DOI :
10.1109/CPEM.2008.4574974