Abstract
This paper describes a measurement set up based on a synchronous sampling system which allows the realization of the henry directly from the ohm and the second. This new set up improved the realization of the henry at METAS by a factor of 5. For the calibration of a 100 mH inductance standard at a frequency of 1 kHz, the expanded uncertainty amounts now to 15 muH/H.
Keywords
calibration; inductance measurement; measurement standards; sampling methods; units (measurement); calibration; frequency 1 kHz; henry; inductance standard; measurement set up; ohm; second; synchronous sampling system; Bridge circuits; Calibration; Electrical resistance measurement; Equations; Impedance; Inductance; Kelvin; Sampling methods; Uncertainty; Voltage;