DocumentCode
2282281
Title
A new architecture for high-accuracy admittance measuring systems
Author
Cabiati, F. ; D´Elia, Vincenzo
Author_Institution
Ist. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
fYear
2002
fDate
16-21 June 2002
Firstpage
178
Lastpage
179
Abstract
The architecture of traditional coaxial systems for highly accurate admittance comparison has been redesigned taking into account the features of active-compensated multi-coaxial connection structures previously developed and investigated. When applied to the conventional two-port transformer bridge, the new approach makes the parasitic cable effects negligible, and allows comparison of four-port standards even in parallel connection.
Keywords
bridge instruments; electric admittance measurement; measurement standards; active compensation; admittance measuring system; coaxial system architecture; four-port standard; multi-coaxial connection structure; parallel connection; parasitic cable effects; two-port transformer bridge; Admittance measurement; Bridge circuits; Coaxial cables; Coaxial components; Conductors; Detectors; Joining processes; Measurement standards; Power cables; Voltage control;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location
Ottawa, Ontario, Canada
Print_ISBN
0-7803-7242-5
Type
conf
DOI
10.1109/CPEM.2002.1034778
Filename
1034778
Link To Document