• DocumentCode
    2282947
  • Title

    A new transfer device for the NRC Travelling Standard Program

  • Author

    Arseneau, R. ; Sutherland, M. ; Zelle, J.

  • Author_Institution
    Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • fYear
    2002
  • fDate
    16-21 June 2002
  • Firstpage
    260
  • Lastpage
    261
  • Abstract
    The paper describes a measurement system, which is now used as a transfer device for the National Research Council of Canada (NRC) Travelling Standard Program. The system is based on two sampling voltmeters and a computer used for data acquisition and analysis. The equipment is used on-site with an ac source and a new 12-position pulse-frequency comparator to provide traceability for the standards maintained by Canadian electrical utilities and electricity meter manufacturers. The accuracy of the system is better than 30 parts in 10/sup 6/. The repeatability of the errors before and after an on-site calibration is generally better than 15 parts in 10/sup 6/.
  • Keywords
    calibration; comparators (circuits); data acquisition; power meters; transfer standards; voltmeters; AC source; NRC Travelling Standard Program; data acquisition; data analysis; electrical utility; electricity meter; measurement system; on-site calibration; pulse-frequency comparator; sampling voltmeter; transfer device; Calibration; Councils; Data analysis; Legislation; Measurement standards; Sampling methods; Standards organizations; Testing; Voltmeters; Watthour meters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Print_ISBN
    0-7803-7242-5
  • Type

    conf

  • DOI
    10.1109/CPEM.2002.1034821
  • Filename
    1034821