DocumentCode
2283702
Title
Determination of the frequency dependence of capacitance standards with ceramic dielectrics
Author
Klionsky, M.D. ; Semenov, Yu.P. ; Moodley, A.
Author_Institution
Mendeleyev (D.I.) Inst. for Metrol., VNIIM, St. Petersburg, Russia
fYear
2002
fDate
16-21 June 2002
Firstpage
346
Lastpage
347
Abstract
The procedure for determination of the frequency dependence (FD) of ceramic capacitance standards characteristic is developed. It is based on application of air capacitance standards with calculable FD, a special summation device and a step calibration method. Measuring instruments are used that allow to compare the capacitance of single capacitors with nominally equal sum capacitance of their combinations. FD of ceramic capacitance standards are determined in the range 100 pF-1 /spl mu/F at frequencies from 50 Hz to 1 MHz. The measurement uncertainty is 0.01-0.1% (k=2).
Keywords
calibration; capacitance measurement; ceramics; dielectric materials; measurement standards; measurement uncertainty; 100 pF to 1 muF; 50 Hz to 1 MHz; air capacitance standard; capacitance standard; ceramic dielectric; frequency dependence; measurement uncertainty; measuring instrument; step calibration method; summation device; Calibration; Capacitance measurement; Capacitance-voltage characteristics; Capacitors; Ceramics; Dielectric measurements; Frequency dependence; Instruments; Measurement uncertainty; Standards development;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location
Ottawa, Ontario, Canada
Print_ISBN
0-7803-7242-5
Type
conf
DOI
10.1109/CPEM.2002.1034864
Filename
1034864
Link To Document