Abstract
A precision set-up and method for calibrating the division ratios of a resistive divider for DC-voltage in the range 10 V to 1000 V is described. The method requires equipment commonly available in most standard laboratories. Uncertainties better than 3/spl times/10/sup -7/ (k = 2) in the ratio is achieved.
Keywords
calibration; measurement standards; measurement uncertainty; voltage dividers; voltage measurement; 10 to 1000 V; DC voltage divider ratio precision calibration; calibration set-up/method; divider voltage range; measurement uncertainty; resistive divider division ratios; Calibration; Electrical resistance measurement; Instruments; Laboratories; Noise measurement; Optical filters; Optical noise; Uncertainty; Voltage; Voltmeters;