• DocumentCode
    2283901
  • Title

    A precision set-up and method for calibrating a DC voltage divider´s ratios from 10 V to 1000 V

  • Author

    Slinde, H. ; Lind, K.

  • Author_Institution
    Nat. Stand. Dept., Justervesenet, Kjeller, Norway
  • fYear
    2002
  • fDate
    16-21 June 2002
  • Firstpage
    372
  • Lastpage
    373
  • Abstract
    A precision set-up and method for calibrating the division ratios of a resistive divider for DC-voltage in the range 10 V to 1000 V is described. The method requires equipment commonly available in most standard laboratories. Uncertainties better than 3/spl times/10/sup -7/ (k = 2) in the ratio is achieved.
  • Keywords
    calibration; measurement standards; measurement uncertainty; voltage dividers; voltage measurement; 10 to 1000 V; DC voltage divider ratio precision calibration; calibration set-up/method; divider voltage range; measurement uncertainty; resistive divider division ratios; Calibration; Electrical resistance measurement; Instruments; Laboratories; Noise measurement; Optical filters; Optical noise; Uncertainty; Voltage; Voltmeters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Print_ISBN
    0-7803-7242-5
  • Type

    conf

  • DOI
    10.1109/CPEM.2002.1034877
  • Filename
    1034877