Title :
The BIPM laser standards at 633 nm and 532 nm simultaneously linked to the SI second using a femtosecond laser in an optical clock configuration
Author :
Long-Sheng Ma ; Robertsson, L. ; Picard, S. ; Chartier, J.-M. ; Ranka, J.K. ; Windeler, R.S.
Author_Institution :
Bur. Int. des Poids et Mesures, Sevres, France
Abstract :
The study of an optical clock based on an iodine-stabilized Nd:YAG laser has been performed using a femtosecond laser. The optical frequency of the iodine-stabilized He-Ne laser at 633 nm and the Nd:YAG laser at 532 nm have been measured simultaneously with a relative uncertainty below 1/spl times/10/sup -12/.
Keywords :
clocks; laser frequency stability; measurement standards; measurement uncertainty; optical variables measurement; 532 nm; 633 nm; BIPM laser standards; He-Ne; I/sub 2/; SI second standard; YAG:Nd; YAl5O12:Nd; femtosecond laser; iodine-stabilized He-Ne laser; iodine-stabilized Nd:YAG laser; optical clock configuration; optical frequency; relative measurement uncertainty; Atom optics; Clocks; Frequency; Laser noise; Laser stability; Optical noise; Optical pumping; Particle beam optics; Pump lasers; Ultrafast optics;
Conference_Titel :
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location :
Ottawa, Ontario, Canada
Print_ISBN :
0-7803-7242-5
DOI :
10.1109/CPEM.2002.1034902