Abstract
The Josephson arbitrary waveform synthesizer can be used as a precision voltage source for both AC and DC signals. Recent improvements in circuit designs have resulted in output voltages greater than 100 mV so that we can investigate AC metrology measurements in the frequency range from 1 kHz to 50 kHz. In this paper we present experimental measurements of several significant error sources that must be considered in order to create an intrinsic AC voltage standard at these frequencies. These results and the corresponding theoretical analysis allow us to move toward creating a practical AC Josephson voltage standard.
Keywords
error analysis; error correction; measurement standards; superconducting arrays; superconducting junction devices; voltage measurement; waveform generators; 1 to 50 kHz; 100 mV; AC Josephson voltage standard; AC metrology frequency range; AC signals; DC signals; Josephson arbitrary waveform synthesizer; circuit designs; error analysis; error measurements; error sources; intrinsic AC voltage standard; output voltages; precision voltage source; Circuits; Error analysis; Filters; Frequency measurement; Measurement standards; Metrology; NIST; Pulse generation; Synthesizers; Voltage measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 2002. Conference Digest 2002 Conference on
Conference_Location
Ottawa, Ontario, Canada
Print_ISBN
0-7803-7242-5
Type
conf
DOI
10.1109/CPEM.2002.1034904
Filename
1034904
Link To Document