DocumentCode :
2288888
Title :
Intraocular pressure measurement devices using the micro reflected air pressure sensor for the pre-diagnosis of the glaucoma
Author :
Kim, Kyoung Hwan ; Jang, Woong-ki ; Kim, Byeong Hee ; Seo, Young Ho
Author_Institution :
Dept. of Mechatron. Eng., Kangwon Nat. Univ., Chuncheon, South Korea
fYear :
2010
fDate :
17-20 Aug. 2010
Firstpage :
907
Lastpage :
910
Abstract :
This work suggested the novel portable tonometer using the micro reflected air pressure sensor for the pre-diagnosis of glaucoma. Because of glaucoma is slowly progressed and not painful, glaucoma patients require a portable pre-diagnosis system which periodically measures the intraocular pressure (IOP) at home. Conventionally IOP was measured by air-puff tonometer where cornea was deformed by short pulse of air pressure and magnitude of the corneal deformation was measured by optic systems such as combination of laser- and photo-diodes. In this work, the micro reflected air pressure sensor was designed, fabricated and tested in order to measure magnitude of the corneal deformation without optic systems. In experimental study, the artificial eyes which had different internal pressures were fabricated, and the internal pressures were measured by the suggested systems.
Keywords :
eye; medical signal processing; pressure measurement; pressure sensors; glaucoma; intraocular pressure measurement devices; micro reflected air pressure sensor; portable tonometer; pre-diagnosis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nanotechnology (IEEE-NANO), 2010 10th IEEE Conference on
Conference_Location :
Seoul
ISSN :
1944-9399
Print_ISBN :
978-1-4244-7033-4
Electronic_ISBN :
1944-9399
Type :
conf
DOI :
10.1109/NANO.2010.5697993
Filename :
5697993
Link To Document :
بازگشت