Title :
Broadband dielectric spectroscopy calibration using calibration liquids with unknown permittivity
Author :
Liu, Siyuan ; Ocket, I. ; Barmuta, Pawel ; Markovic, T. ; Lewandowski, Andreas ; Schreurs, Dominique ; Nauwelaers, B.
Author_Institution :
Div. ESAT-TELEMIC, Univ. of Leuven, Leuven, Belgium
Abstract :
This paper presents a calibration method for broadband dielectric spectroscopy of micro-liter samples of biological liquids with transmission line sensors up to 110 GHz. The method uses water and methanol as calibration liquids to perform a probe-tip-to-liquid calibration without prior knowledge of their complex permittivity. Extracted complex permittivities of the calibration liquids show good agreement with literature values. Furthermore, sensors with different first-tier calibrations and geometries show consistent results.
Keywords :
biological techniques; calibration; microfluidics; millimetre wave detectors; millimetre wave measurement; organic compounds; permittivity measurement; biological liquids; broadband dielectric spectroscopy calibration method; calibration liquids; complex permittivity extraction; first tier calibration; methanol; microliter samples; probe-tip-to-liquid calibration; transmission line sensors; unknown permittivity; water; Calibration; Coplanar waveguides; Liquids; Permittivity; Scattering parameters; Sensors; Standards; biology; calibration; dielectrics; millimeter wave; on-wafer; spectroscopy;
Conference_Titel :
Microwave Measurement Conference (ARFTG), 2014 84th ARFTG
Conference_Location :
Boulder, CO
DOI :
10.1109/ARFTG.2014.7013421