An on-line monitoring technique for electrode degradation in bio-fluidic microsystems
Author :
Al-Gayem, Q. ; Liu, H. ; Richardson, A. ; Burd, N. ; Kumar, M.
Author_Institution :
Centre for Microsyst. Eng., Lancaster Univ., Lancaster, UK
fYear :
2010
fDate :
2-4 Nov. 2010
Firstpage :
1
Lastpage :
10
Abstract :
This paper presents a solution for detecting degradation in electrodes that interface to fluidic or biological systems that forms the basis of numerous actuation and sensing mechanisms in the bio-fluidics field. In this solution, a mid-frequency oscillation test strategy is proposed and evaluated experimentally on an array of electrodes. This technique is based on the sensitivity of the bio-fluidic interface capacitance to degradation, contamination and fouling.
Keywords :
electrodes; microfluidics; bio-fluidic microsystem; contamination; electrode degradation; fouling; mid-frequency oscillation test strategy; online monitoring technique;