• DocumentCode
    2311293
  • Title

    Surface biopotential monitoring by needle type micro electrode array

  • Author

    Pan, Li-Chern ; Lin, Po-Wei ; Tseng, Fan-Gang ; Lin, Chun

  • Author_Institution
    Graduate Inst. of Oral Rehabilitation Sci., Taipei Med. Univ., Taiwan
  • Volume
    1
  • fYear
    2002
  • fDate
    2002
  • Firstpage
    221
  • Abstract
    The current study reports a novel bioelectrode design based on a series of micro needle arrays for the measurement of surface biopotential. Without the pain sensation, the goal of these micro needles is to penetrate skin and minimize the variations in skin impedance, as well as body motion. Needle electrodes with high depth definition were fabricated and tested on a de-haired rat skin surface for impedance measurements. Experiment results showed a prominent reduction in electrode impedance when compared with traditional silver/silver-chloride electrodes. A quasi-linear relationship was also observed between electrode surface area and measured skin/electrode impedance.
  • Keywords
    arrays; bioelectric potentials; biomedical electrodes; biomedical measurement; microelectrodes; patient monitoring; skin; bioelectrode design; high depth definition; impedance measurements; micro needle array; needle type micro electrode array; skin impedance; surface biopotential monitoring; Current measurement; Electrodes; Impedance measurement; Monitoring; Needles; Pain; Silver; Skin; Surface impedance; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2002. Proceedings of IEEE
  • Print_ISBN
    0-7803-7454-1
  • Type

    conf

  • DOI
    10.1109/ICSENS.2002.1037086
  • Filename
    1037086