DocumentCode
2311293
Title
Surface biopotential monitoring by needle type micro electrode array
Author
Pan, Li-Chern ; Lin, Po-Wei ; Tseng, Fan-Gang ; Lin, Chun
Author_Institution
Graduate Inst. of Oral Rehabilitation Sci., Taipei Med. Univ., Taiwan
Volume
1
fYear
2002
fDate
2002
Firstpage
221
Abstract
The current study reports a novel bioelectrode design based on a series of micro needle arrays for the measurement of surface biopotential. Without the pain sensation, the goal of these micro needles is to penetrate skin and minimize the variations in skin impedance, as well as body motion. Needle electrodes with high depth definition were fabricated and tested on a de-haired rat skin surface for impedance measurements. Experiment results showed a prominent reduction in electrode impedance when compared with traditional silver/silver-chloride electrodes. A quasi-linear relationship was also observed between electrode surface area and measured skin/electrode impedance.
Keywords
arrays; bioelectric potentials; biomedical electrodes; biomedical measurement; microelectrodes; patient monitoring; skin; bioelectrode design; high depth definition; impedance measurements; micro needle array; needle type micro electrode array; skin impedance; surface biopotential monitoring; Current measurement; Electrodes; Impedance measurement; Monitoring; Needles; Pain; Silver; Skin; Surface impedance; Testing;