Research on the sensor calibration of pressure transmitter based on the least square method
Author
Zhiping Wang ; Aidong Xu ; Kai Wang ; Yan Song ; Bingjun Yan
Author_Institution
Shenyang Inst. of Autom., Shenyang, China
fYear
2014
fDate
19-23 Oct. 2014
Firstpage
389
Lastpage
392
Abstract
The accuracy of capacitance pressure sensor is easily disturbed by work environment. In order to improve the accuracy of pressure sensor, the paper researches on the sensor calibration of pressure transmitter based on the least square method. The measurement circuit is designed to convert the output of sensor from the capacitance value to the cycle value. Working quadric surface is fitted with the value of environmental temperature, the output of pressure sensor and setting pressure by the least square method. The pressure transmitters can accurately measure at different temperatures by the experiment.
Keywords
calibration; capacitive sensors; least squares approximations; pressure sensors; capacitance pressure sensor; least square method; measurement accuracy; measurement circuit; pressure transmitter; sensor calibration; working quadric surface; Abstracts; Accuracy; Instruments; Pressure measurement; Transmitters; pressure transmitter; sensor calibration; the least square method;
fLanguage
English
Publisher
ieee
Conference_Titel
Signal Processing (ICSP), 2014 12th International Conference on
Conference_Location
Hangzhou
ISSN
2164-5221
Print_ISBN
978-1-4799-2188-1
Type
conf
DOI
10.1109/ICOSP.2014.7015034