DocumentCode
2317047
Title
Adaptive threshold QRS detector with best channel selection based on a noise rating system
Author
Chiarugi, F. ; Sakkalis, V. ; Emmanouilidou, D. ; Krontiris, T. ; Varanini, M. ; Tollis, I.
Author_Institution
Inst. of Comput. Sci., FORTH, Heraklion
fYear
2007
fDate
Sept. 30 2007-Oct. 3 2007
Firstpage
157
Lastpage
160
Abstract
QRS detection performance can depend on the type of noise present in each lead involved in the overall processing. A common approach to QRS detection is based on a QRS enhanced signal obtained from the derivatives of the pre-filtered leads. However, the signal pre-filtering cannot be able to perform a complete noise rejection and the use of derivatives can enhance the noise as well. In many cases the noise occurs only on one lead and the addition of a noisy lead to the QRS enhanced signal decreases the overall detection performances of the QRS detector. For this reason the noise estimation on each channel, providing information for the channel inclusion or rejection in building the QRS enhanced signal, can improve the overall performances of the QRS detector. The results have been evaluated on the 48 records of the MIT-BIH Arrhythmia Database where each ECG record is composed by 2 leads sampled at 360 Hz for a total duration of about 30 minutes. The annotated QRSs are 109494 in total. The results have been very satisfying on all the annotated QRSs and, with the inclusion of an automatic criterion for ventricular flutter detection, a sensitivity=99.76% and a positive predictive value=99.81% have been obtained.
Keywords
electrocardiography; filtering theory; medical information systems; medical signal detection; medical signal processing; ECG signal recording; MIT-BIH arrhythmia database; QRS detection algorithm; QRS enhanced signal; adaptive threshold QRS detector; electrocardiogram; frequency 360 Hz; noise rating system; prefiltered signal; signal noise estimation; time 30 min; ventricular flutter detection; Band pass filters; Cardiology; Computer science; Databases; Detectors; Electrocardiography; Frequency; Noise level; Noise reduction; Nonlinear filters;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2007
Conference_Location
Durham, NC
ISSN
0276-6547
Print_ISBN
978-1-4244-2533-4
Electronic_ISBN
0276-6547
Type
conf
DOI
10.1109/CIC.2007.4745445
Filename
4745445
Link To Document