DocumentCode :
2318336
Title :
Rectangular-waveguide vector-network-analyzer calibrations with imperfect test ports
Author :
Williams, Dylan F.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
fYear :
2010
fDate :
Nov. 30 2010-Dec. 3 2010
Firstpage :
1
Lastpage :
8
Abstract :
We present a strategy for correcting for imperfect interfaces between the test ports of a vector network analyzer, the calibration standards and the devices under test. This corrects for the inconsistencies in calibrations introduced by use of flush thrus and flat shorts as calibration standards. The approach is based on equivalent standard definitions that are easy to implement in conventional network analyzers. We present analytic formulas for these definitions and demonstrate them in WR-90 rectangular waveguide.
Keywords :
calibration; network analysers; rectangular waveguides; WR-90 rectangular waveguide; device under test; imperfect interface correction; imperfect test ports; vector-network-analyzer calibrations; Calibration; submillimeter wave; vector network analyzer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Measurement Symposium (ARFTG), 2010 76th ARFTG
Conference_Location :
Clearwater Beach, FL
Print_ISBN :
978-1-4244-7447-9
Type :
conf
DOI :
10.1109/ARFTG76.2010.5700048
Filename :
5700048
Link To Document :
بازگشت