DocumentCode :
2338329
Title :
Parallel resonance Q meter and its applications
Author :
Kataoka, Teruyuki ; Kanno, Makoto
Author_Institution :
Dept. of Electron. Eng., Tamagawa Univ., Tokyo, Japan
fYear :
1994
fDate :
10-12 May 1994
Firstpage :
1008
Abstract :
Principle of a parallel resonance Q meter in which the measuring capacitor is connected as a three-terminal element to eliminate the effects of the stray capacitances, is introduced. Experimental results with small measuring capacitances (about 100 pF) shows that the stray capacitances about 10 times as much as the measuring capacitance is eliminated. As an application of this Q meter, an LC oscillator is introduced, which generates a digital signal, converting the capacitance to a frequency
Keywords :
Q-factor measurement; capacitance measurement; coaxial cables; convertors; meters; LC oscillator; capacitance frequency convertor; digital signal; parallel resonance Q meter; stray capacitances; three-terminal element; Capacitance measurement; Capacitors; Frequency measurement; Impedance; Joining processes; Oscillators; Q measurement; RLC circuits; Resonance; Resonant frequency;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1994. IMTC/94. Conference Proceedings. 10th Anniversary. Advanced Technologies in I & M., 1994 IEEE
Conference_Location :
Hamamatsu
Print_ISBN :
0-7803-1880-3
Type :
conf
DOI :
10.1109/IMTC.1994.351934
Filename :
351934
Link To Document :
بازگشت