    Study on stabilized zero-point of electromagnetic flowmeter with rapid excitation

  • Author

    Saito, OK ; Sakurai, Y. ; Okayama, T.

  • Author_Institution
    Instrum. Div., Hitachi Ltd., Ibaraki, Japan
  • fYear
    1994
  • fDate
    10-12 May 1994
  • Firstpage
    829
  • Abstract
    This paper describes a designing method to prevent zero-point drift which are susceptible to result when a rapid excitation is employed with electromagnetic flowmeter. It is shown how to prevent adverse effect of magnetic flux differential noise in rapid excitation, the new items of the composite alternated constant current excitation circuit, magnetic circuit configuration, and drift-corrected signal processing. As a result, zero-point drift which were observed ±2% in the maximum sensitivity range of 0.1 m/s have been reduced to ±0.15% or less
  • Keywords
    electromagnetic devices; electromagnetic interference; exciters; flow measurement; flowmeters; interference suppression; random noise; signal processing; composite alternated constant current excitation circuit; drift-corrected signal processing; electromagnetic flowmeter; industrial measurements; magnetic circuit configuration; magnetic flux differential noise; process measurements; rapid excitation; stabilized zero-point; zero-point drift; Circuits; Coils; DC-DC power converters; Electromagnetic measurements; Frequency; Instruments; Magnetic flux; Magnetic noise; Switches; Voltage control;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1994. IMTC/94. Conference Proceedings. 10th Anniversary. Advanced Technologies in I & M., 1994 IEEE
  • Conference_Location
    Hamamatsu
  • Print_ISBN
    0-7803-1880-3
  • Type

    conf

  • DOI
    10.1109/IMTC.1994.351981
  • Filename
    351981