Title :
Comparison Between Thermoelectric and Bolometric Microwave Power Standards
Author :
Brunetti, L. ; Oberto, L. ; Sellone, M. ; Vremera, E.T.
Author_Institution :
Electromagn. Div., Ist. Naz. di Ricerca Metrol., Turin, Italy
Abstract :
This paper describes a comparison of microwave power standards based on thermoelectric sensors to an analogous standard based on bolometric sensors. Measurements have been carried out with the classical twin-type microcalorimeter, fitted with N-connector test ports suitable for the frequency band 0.05-18 GHz. An appropriate measurand definition is given as being suitable for both standard types. A system accuracy assessment is performed applying the Gaussian error propagation through the mathematical models that interpret the microcalorimeter response in each case. The results highlight advantages and weaknesses of each power standard type.
Keywords :
Gaussian noise; bolometers; calorimeters; measurement errors; measurement standards; microsensors; microwave measurement; temperature sensors; thermoelectric devices; Gaussian error propagation; N-connector test port; analogous standard; bolometric sensor; frequency 0.05 GHz to 18 GHz; mathematical model; microwave power standard; system accuracy assessment; thermoelectric sensor; twin type microcalorimeter; Broadband microcalorimeter; microwave measurements; microwave standards; power measurement; thermistors; thermoelectric devices;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2013.2253914