• DocumentCode
    2348090
  • Title

    2C-1 Contrast Resonance Imaging with Embedded Ultrasound Imaging Pulses

  • Author

    Shi, William T. ; Hall, Christopher S. ; Rafter, Patrick G.

  • Author_Institution
    Philips Res. North America, Briarcliff Manor, NY
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    212
  • Lastpage
    215
  • Abstract
    A new ultrasound contrast imaging modality called Contrast Resonance Imaging (CRI) is proposed, which is based on our finding that, for a short imaging pulse, the nonlinear scattering signal from resonant contrast microbubbles can be greatly enhanced if the imaging pulse is embedded at the end of a relatively long pulse. The new imaging technique is evaluated with in vitro laboratory A-line measurements, achieving a contrast-to-tissue ratio of approximate 40 dB. For a proof of the concept, the new imaging mode is implemented on an iE33 ultrasound scanner and images of a contrast flow phantom with robust contrast signals are obtained
  • Keywords
    acoustic resonance; bubbles; ultrasonic imaging; ultrasonic scattering; CRI; contrast flow phantom; contrast resonance imaging; contrast-tissue ratio; embedded ultrasound imaging pulses; iE33 ultrasound scanner; imaging pulse; in vitro laboratory A-line measurements; nonlinear scattering signal; resonant contrast microbubbles; Frequency; Imaging phantoms; In vitro; Laboratories; Numerical simulation; Particle scattering; Pulse modulation; Resonance; Ultrasonic imaging; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.66
  • Filename
    4151922