• DocumentCode
    2348552
  • Title

    2D-2 Automatic Detection of Ultrasound Contrast Microbubble Shell Rupture

  • Author

    Ammi, A.Y. ; Bridal, L. ; Mamou, J. ; Wang, G.I. ; O´Brien, W.D.

  • Author_Institution
    Lab. d´´Imagerie Parametrique, Univ. Paris
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    297
  • Lastpage
    300
  • Abstract
    Characterizing occurrence of ultrasound contrast agent (UCA) microbubble destruction is important for development of functional and therapeutic applications. Previously [Ammi, et. al., 2006], it was demonstrated that post-excitation acoustic emissions detected with passive cavitation detection (PCD) result from inertial cavitation (IC) of UCA after shell rupture. That work relied on time-consuming visual inspection of PCD data to identify IC signals and characterized only minimum rarefactional pressure thresholds for rupture of a single Optisontrade microbubble in the sampled population. This work introduces an algorithm for automatic detection of IC signals. The algorithm was applied to the 71424 waveforms in the PCD data set. At each incident frequency (0.9, 2.8 or 4.6 MHz) and pulse duration (3, 5 or 7 cycles) combination, ruptured microbubble occurrence with incident peak rarefactional pressure (PRP) was well described by a logistic regression curve with an inflection point near 50% occurrence and a plateau at 100%. With a 5-cycle pulse duration at 0.9, 2.8 and 4.6 MHz the incident PRPs leading to 5% microbubble rupture were 0.66, 0.83 and 1.1 MPa; and 1.1, 1.6 and 2.5 MPa for 50%. This automatic algorithm combined with the PCD approach provides a practical tool for the characterization of UCA destruction occurrence across a significant range of incident PRPs
  • Keywords
    biomedical ultrasonics; bubbles; cavitation; drugs; fracture; 0.66 MPa; 0.83 MPa; 0.9 MHz; 1.1 MPa; 1.6 MPa; 2.5 MPa; 2.8 MHz; 4.6 MHz; Optisontrade; microbubble; passive cavitation detection; peak rarefactional pressure; shell rupture; ultrasound contrast agent; Acoustic emission; Acoustic measurements; Acoustic pulses; Acoustic signal detection; Biomedical optical imaging; Frequency; Integrated circuit noise; Signal detection; Signal processing; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.87
  • Filename
    4151943