    1H-4 Use of Acoustic Scattering to Monitor Kidney Stone Fragmentation During Shock Wave Lithotripsy

  • Author

    Owen, Neil R. ; Sapozhnikov, Oleg A. ; Bailey, Michael R. ; Trusov, Leonid ; Crum, Lawrence A.

  • Author_Institution
    Appl. Phys. Lab., Washington Univ., Seattle, WA
  • fYear
    2006
  • fDate
    2-6 Oct. 2006
  • Firstpage
    736
  • Lastpage
    739
  • Abstract
    It is currently difficult to assess whether a kidney stone has fractured during shock wave lithotripsy. Here we report the calculation and measurement of shock wave scattering by stone models in water. Calculations were based on linear elastic theory to find pressure in the fluid and stress in the stone models, and on scattering theory to find radiation from the stone models. Measurements were made with a spherical, broadband receiver. Calculation and measurement agree well in the time domain and through frequency analysis of detected acoustic scattering it was possible to distinguish between fractured and intact model stones. Cavitation was visualized with high speed photography and was not a dominant effect in the measurements
  • Keywords
    biomedical ultrasonics; cavitation; kidney; shock waves; ultrasonic scattering; acoustic scattering; cavitation; high speed photography; kidney stone fragmentation; linear elastic theory; shock wave lithotripsy; Acoustic measurements; Acoustic scattering; Frequency domain analysis; Frequency measurement; Lithotripsy; Monitoring; Shock waves; Stress; Time domain analysis; Time measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2006. IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1051-0117
  • Print_ISBN
    1-4244-0201-8
  • Electronic_ISBN
    1051-0117
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2006.197
  • Filename
    4152053