DocumentCode :
2350868
Title :
5H-4 An Implantable Ultrasonic Doppler Blood Flowmeter
Author :
Cheng, Xiaoyang ; Chen, Jingkuang ; Zhang, Tongsheng ; Chen, Chien-Chang ; Cheng, Yu-Ting ; Wang, Mengli
Author_Institution :
Dept. of Electr. & Comput. Eng.,, New Mexico Univ., Albuquerque, NM
fYear :
2006
fDate :
2-6 Oct. 2006
Firstpage :
808
Lastpage :
811
Abstract :
A miniature ultrasonic transducer array for minimally invasive blood flow measurement has been developed. Compare with laser Doppler flowmeter, this tool is much smaller in size and is capable of reaching deep inside the tissue to monitor blood flow not accessible by the laser beam. Compared with traditional ultrasound Doppler flowmeter, this miniature tool is able to achieve more accurate measurement by using high frequency ultrasounds. The silicon substrate of the transducers array was micromachined into a probe shape using double-sided dry etching. The typical cross-sectional dimension of the silicon probe is 40 mum times 80 mum. Arrays comprising 24-32 transducers were prototyped and tested. The 46 mum-diameter drum shaped ultrasound transducers operate at a voltage lower than 40 volts, generate and receive ultrasound at a central frequency of 5.8 MHz, and has a bandwidth of about 80%. A new poly silicon surface micromachining process was developed for fabrication of the transducer. The new process separates all the conductors from the silicon nitride film and reduces the charging problem associated with nitride-conductor interface
Keywords :
Doppler measurement; biomedical transducers; biomedical ultrasonics; blood flow measurement; micromachining; micromechanical devices; silicon; ultrasonic transducer arrays; 46 micron; 5.8 MHz; double-sided dry etching; implantable ultrasonic Doppler blood flowmeter; miniature ultrasonic transducer array; polysilicon surface micromachining process; silicon substrate; ultrasound transducers; Blood flow; Conductive films; Frequency; Minimally invasive surgery; Probes; Silicon; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers; Ultrasonic variables measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2006. IEEE
Conference_Location :
Vancouver, BC
ISSN :
1051-0117
Print_ISBN :
1-4244-0201-8
Electronic_ISBN :
1051-0117
Type :
conf
DOI :
10.1109/ULTSYM.2006.211
Filename :
4152067
Link To Document :
بازگشت