Orthopaedic alloy electrochemical behavior: the case of Ti-6Al-7Nb
Author :
Lewis, Gladius ; Shaw, Kevin M.
Author_Institution :
Dept. of Mech. Eng., Memphis Univ., TN, USA
fYear :
1995
fDate :
7-9 Apr 1995
Firstpage :
235
Lastpage :
238
Abstract :
The static electrochemical behavior of Ti-6Al-7Nb alloy specimens in three lactated Ringer´s solution-based electrolytes was investigated using the d.c. polarization method. It was found that, relative to the case with Ringer´s solution having pH = 6.25, decreasing the solution pH to 1 increases the corrosion rate of the alloy but adding bovine serum to the solution only marginally affects the rate. A critical appraisal of the polarization method is presented, and comments on the clinical usefulness of the results are included
Keywords :
aluminium alloys; bone; corrosion; niobium alloys; prosthetics; titanium alloys; DC polarization method; Ti-6Al-7Nb alloy specimens; Ti-Al-Nb; biocompatibility; biostability; bovine serum; corrosion rate; endoprostheses; lactated Ringer´s solution-based electrolytes; orthopaedic alloy electrochemical behavior; polarization method; solution pH; static electrochemical behavior; Bovine; Computer aided software engineering; Corrosion; Implants; In vivo; Iron alloys; Orthopedic surgery; Polarization; Prosthetics; Steel;