• DocumentCode
    2358116
  • Title

    Heart rate detection in highly noisy handgrip electrocardiogram

  • Author

    Lin, C.C. ; Hu, W.C. ; Chen, CM ; Weng, CH

  • Author_Institution
    Dept. of Electr. Eng., Nat. Chin-Yi Univ. of Technol., Taiping
  • fYear
    2008
  • fDate
    14-17 Sept. 2008
  • Firstpage
    477
  • Lastpage
    480
  • Abstract
    This study is to develop the handgrip heart rate detection methodology based on template matching algorithm. A sixth-order Butterworth bandpass filter with a passband of 1 to 30 Hz was used to filter out baseline shift and noise above 30 Hz including power line noise. The resulting QRS waveform will be processed by the matching filter and ready for heart beat detection. The result of detected heart beat will be processed by a weighting function to further minimize the false heartbeat detection. The performance of heart rate detection algorithm was tested under resting and exercising condition. Five Taiwanese male subjects averaged age of 25 were tested. The experimental results showed that the error rates were 0.78% (total of 766 heart beats) at resting condition and 3.83% (total of 3077 heart beats) during exercise.
  • Keywords
    electrocardiography; matched filters; medical signal detection; medical signal processing; QRS waveform; Taiwanese male subjects; baseline shift; exercise; frequency 1 Hz to 30 Hz; heart rate detection; highly noisy handgrip electrocardiogram; matching filter; power line noise; resting condition; sixth-order Butterworth bandpass filter; template matching algorithm; weighting function; Band pass filters; Electrocardiography; Electrodes; Heart beat; Heart rate; Heart rate detection; Matched filters; Noise level; Semiconductor device noise; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2008
  • Conference_Location
    Bologna
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4244-3706-1
  • Type

    conf

  • DOI
    10.1109/CIC.2008.4749082
  • Filename
    4749082