Heart rate detection in highly noisy handgrip electrocardiogram
Author
Lin, C.C. ; Hu, W.C. ; Chen, CM ; Weng, CH
Author_Institution
Dept. of Electr. Eng., Nat. Chin-Yi Univ. of Technol., Taiping
fYear
2008
fDate
14-17 Sept. 2008
Firstpage
477
Lastpage
480
Abstract
This study is to develop the handgrip heart rate detection methodology based on template matching algorithm. A sixth-order Butterworth bandpass filter with a passband of 1 to 30 Hz was used to filter out baseline shift and noise above 30 Hz including power line noise. The resulting QRS waveform will be processed by the matching filter and ready for heart beat detection. The result of detected heart beat will be processed by a weighting function to further minimize the false heartbeat detection. The performance of heart rate detection algorithm was tested under resting and exercising condition. Five Taiwanese male subjects averaged age of 25 were tested. The experimental results showed that the error rates were 0.78% (total of 766 heart beats) at resting condition and 3.83% (total of 3077 heart beats) during exercise.
Keywords
electrocardiography; matched filters; medical signal detection; medical signal processing; QRS waveform; Taiwanese male subjects; baseline shift; exercise; frequency 1 Hz to 30 Hz; heart rate detection; highly noisy handgrip electrocardiogram; matching filter; power line noise; resting condition; sixth-order Butterworth bandpass filter; template matching algorithm; weighting function; Band pass filters; Electrocardiography; Electrodes; Heart beat; Heart rate; Heart rate detection; Matched filters; Noise level; Semiconductor device noise; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2008
Conference_Location
Bologna
ISSN
0276-6547
Print_ISBN
978-1-4244-3706-1
Type
conf
DOI
10.1109/CIC.2008.4749082
Filename
4749082
Link To Document