• DocumentCode
    235959
  • Title

    Calibrations for resistance welding at high pulsed currents

  • Author

    Djokic, Branislav

  • Author_Institution
    Meas. Sci. & Stand, Nat. Res. Council of Canada, Ottawa, ON, Canada
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    12
  • Lastpage
    13
  • Abstract
    Rogowski coils are widely used as high pulsed current sensors in resistance welding. Their operating conditions reflect on requirements for their calibration. A new calibration system has been developed at NRC Canada for calibrating current sensing coils and weld current monitors for AC resistance welding. The calibration system, which includes a high current source and a coaxial copper cage, has been evaluated at peak pulsed currents of up to 80 kA, at 60 Hz. The best expanded measurement uncertainties (k=2) have been estimated to be within 250 μA/A for magnitude and 175 μrad for phase.
  • Keywords
    calibration; coils; constant current sources; electric current measurement; electric resistance measurement; electric sensing devices; measurement uncertainty; phase estimation; resistance welding; AC resistance welding; NRC Canada; Rogowski coil; calibration system; coaxial copper cage; current source; frequency 60 Hz; high pulsed current sensor coil; measurement uncertainty; weld current monitoring; Calibration; Coils; Current measurement; Measurement uncertainty; Monitoring; Resistance; Welding; AC current measurement; Rogowski coils; calibrations; digital sampling; measurement uncertainty; pulsed currents; weld current monitors/meters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898234
  • Filename
    6898234