• DocumentCode
    236025
  • Title

    Calibration of capacitance in the frequency range up to 1 MHz

  • Author

    Dan Bee Kim ; Hyung-Kew Lee ; Wan-Seop Kim

  • Author_Institution
    Korea Res. Inst. of Sci. & Technol., Daejeon, South Korea
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    90
  • Lastpage
    91
  • Abstract
    This paper describes two different ways to calibrate the capacitance in the high frequency range up to 13 MHz. First, a dual source bridge was configured using two voltage sources and then the ratio of two capacitors was measured in the frequency range from 1 kHz to 1 MHz. The second method known as the Z-matrix method, allowed us to measure the capacitance value up to 13 MHz. The measured values were compared with the reference values calibrated by the PTB and Agilent Technologies, respectively, and they were also compared with each other. The cross-checks showed that both methods are reliable to an order of 10-5 at 1 MHz.
  • Keywords
    bridge circuits; calibration; capacitance measurement; capacitors; Agilent Technology; PTB; Z-matrix method; calibration; capacitance measurement; capacitor; dual source bridge; voltage source; Bridge circuits; Calibration; Capacitance; Capacitance measurement; Capacitors; Frequency measurement; Standards; Capacitance measurement; Z-matrix method; dual source bridge; high frequency;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898273
  • Filename
    6898273