Title :
Building-up the DC voltage scale for sources and meters
Author :
Early, Murray D. ; Christian, Laurie A. ; Lawson, Tim B.
Author_Institution :
Meas. Stand. Lab. of New Zealand (MSL), Lower Hutt, New Zealand
Abstract :
Previous work has shown that an automated build-up method, called the Reference Step Method (RSM), can measure the gain ratios of a dc voltage source with an uncertainty of 0.2 μV/V up to 1000 V. We have extended this method to also establish the gain ratio of the digital voltmeter (DVM) used in the build-up method. This only requires a few additional measurements so that the gain ratios of the source and meter can be measured simultaneously. Initial measurements suggest that comparable uncertainties can also be achieved for the DVM.
Keywords :
digital voltmeters; electric current measurement; gain measurement; voltage measurement; DC voltage source scale; DVM; RSM; build-up method; digital voltmeter; gain ratio measurement; reference step method; Gain measurement; Linearity; Measurement uncertainty; Meter reading; Standards; Uncertainty; Voltage measurement; dc build-up; dc range calibration; dc voltage ratio; dc voltage scale; measurement techniques;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898297