• DocumentCode
    236187
  • Title

    10 V programmable Josephson voltage standard and its application in direct comparison with conventional Josephson voltage standard

  • Author

    Yi-hua Tang ; Sims, June E.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    256
  • Lastpage
    257
  • Abstract
    This paper briefly describes the working principle of the 10 V programmable Josephson voltage standard (PJVS) that was developed at the National Institute of Standards and Technology (NIST) and how to use it in a direct comparison with a conventional Josephson voltage standard (JVS). Manual and automatic comparison methods were investigated to verify the agreement between the two types of Josephson standards. The 10 V PJVS will be used as a transfer standard in the 2014 JVS Interlaboratory Comparison (ILC) that is organized by the National Conference of Standard Laboratories International (NCSLI).
  • Keywords
    Josephson effect; measurement standards; voltage measurement; automatic comparison; manual comparison; programmable Josephson voltage standard; transfer standard; voltage 10 V; 1f noise; Floors; NIST; Protocols; Uncertainty; Voltage measurement; automatic comparison; conventional Josephson voltage standard; programmable Josephson voltage standard; uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898356
  • Filename
    6898356