• DocumentCode
    236260
  • Title

    Uncertainty analysis of non-coherent sampling phase meter with four parameter sine wave fitting by means of Monte Carlo

  • Author

    Sira, Martin ; Maslan, Stanislav

  • Author_Institution
    Czech Metrol. Inst., Brno, Czech Republic
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    334
  • Lastpage
    335
  • Abstract
    Uncertainty analysis of a very simple sampling phase meter based on four parameter sine wave fitting is presented in this paper. An uncertainty of a sample time and a measured voltage is assigned to every sampled point and a total uncertainty of a phase difference is calculated by means of the Monte Carlo method. Results show the total uncertainty is lower than 0.5 millidegree for selected digitizer. A noise of the measured signal or the digitizer is the largest contribution to the total uncertainty.
  • Keywords
    Monte Carlo methods; electric noise measurement; measurement uncertainty; phase meters; phase noise; voltage measurement; Monte Carlo method; noncoherent sampling phase meter uncertainty analysis; parameter sine wave fitting; phase difference uncertainty; signal noise measurement; voltage measurement; Barium; Generators; Measurement uncertainty; Monte Carlo methods; Phase measurement; Uncertainty; Voltage measurement; Monte Carlo; measurement uncertainty; phase difference; sine wave fitting;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898395
  • Filename