• DocumentCode
    236273
  • Title

    Linearity analysis of a calorimetric thermal voltage converter for RF-DC transfer difference up to 1 GHz

  • Author

    Yueyan Shan ; Yu Song Meng ; Filipski, Piotr S.

  • Author_Institution
    Nat. Metrol. Centre, Agency for Sci., Technol. & Res. (A*STAR), Singapore, Singapore
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    346
  • Lastpage
    347
  • Abstract
    Linearity analysis of a calorimetric thermal voltage converter (CTVC) designed by the National Research Council of Canada for RF-DC transfer difference up to 1 GHz is presented in this paper. At higher frequencies, the CTVC is evaluated in terms of the calibration factor of a RF power sensor and the reflection coefficients at its Type-N connector. Good performance has been observed which indicates that CTVC could be a potential candidate for calibrating RF power sensors in terms of its RF-DC transfer difference.
  • Keywords
    calibration; calorimetry; electric connectors; electric sensing devices; transfer standards; voltage measurement; CTVC; RF power sensor calibration; RF-DC transfer difference; Type-N connector; calorimetric thermal voltage converter; frequency 1 GHz; linearity analysis; reflection coefficients; Calibration; Connectors; Linearity; Power measurement; Radio frequency; Reflection coefficient; Standards; CTVC; RF-DC transfer difference; linearity; mathematical modeling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898401
  • Filename
    6898401