• DocumentCode
    236402
  • Title

    A measurement chain for the determination of RK using a calculable capacitor

  • Author

    Fletcher, Nick ; Goebel, R.

  • Author_Institution
    BIPM, Sevres, France
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    474
  • Lastpage
    475
  • Abstract
    The BIPM has an ongoing project to establish a Thompson Lampard calculable capacitor for use in the determination of the von Klitzing constant, RK. This paper describes key parts of the measurement chain used to provide the link between the dc quantum Hall effect and the calculable capacitor. The bridge used to calibrate a conventional 1 pF standard against a known 0.4 pF capacitance change is presented. The target relative uncertainty for the overall measurement is 1 part in 108.
  • Keywords
    capacitance measurement; capacitors; constants; measurement standards; measurement uncertainty; quantum Hall effect; BIPM; DC quantum Hall effect; RK determination; Thompson Lampard calculable capacitor; calibration; capacitance measurement standard; relative uncertainty; von Klitzing constant determination; Bridge circuits; Capacitance measurement; Capacitors; Quantum capacitance; Silicon; Standards; Capacitance; Coaxial Bridges; Fundamental Constants; Thompson Lampard;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898465
  • Filename
    6898465