Title :
Update from the BIPM watt balance
Author :
Fang, Haiyang ; Kiss, A. ; Lavergne, T. ; Robertsson, L. ; de Mirandes, E. ; Solve, S. ; Picard, A. ; Stock, Michiel
Author_Institution :
Bur. Int. des Poids et Mesures (BIPM), Sevres, France
Abstract :
This paper reports the latest progress made on the BIPM watt balance project. The type A uncertainty of the Planck constant measurements was reduced from 1.4 parts in 106 to about 5 parts in 107. The new magnet was successively fabricated and assembled with an expected high uniformity of the flux density of several parts in 104. Several new components are being introduced into the apparatus which we expect to reduce the type B measurement uncertainty significantly.
Keywords :
assembling; balances; magnetic flux; magnets; measurement uncertainty; wattmeters; BIPM watt balance project; Planck constant measurement uncertainty; assembling; flux density uniformity; magnet; type B measurement uncertainty; Coils; Force; Magnetic field measurement; Magnetic flux; Standards; Uncertainty; Vibrations; Interferometry; Kilogram; Mass metrology; Planck constant; Watt Balance;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-4799-5205-2
DOI :
10.1109/CPEM.2014.6898583