• DocumentCode
    236631
  • Title

    Linearity calibration of wideband digitizer using binary inductive voltage divider

  • Author

    Jian Feng ; Leibing Shi ; Yi Qin ; Feng Xu

  • Author_Institution
    Shanghai Inst. of Meas. & Testing Technol., Shanghai, China
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    726
  • Lastpage
    727
  • Abstract
    Self-calibration theory and the guarding method for binary inductive voltage divider(IVD) is described. The linearity of binary IVD is in level of 10-7 at frequency of 10 kHz and 10-6 at frequency of 100 kHz. Method to measure linearity of wideband digitizer using binary IVD is introduced. Nonlinearity is no more than 20 μV V at 1 kHz and 10 kHz, and 30 μV V at 100 kHz for full 10-Vpp scale. And the similar linearity curves are found at three frequencies.
  • Keywords
    analogue-digital conversion; calibration; voltage dividers; binary inductive voltage divider; frequency 1 kHz; frequency 10 kHz; frequency 100 kHz; guarding method; linearity calibration; self-calibration theory; wideband digitizer; Amplifiers; Calibration; Connectors; Linearity; Standards; Voltage measurement; Wideband; Binary IVD; digitizer; linearity; self-calibration; wideband;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898591
  • Filename
    6898591