• DocumentCode
    236633
  • Title

    Calibration methods of a CC and BCG for a high-accuracy AC current ratio standard up to 4 kHz

  • Author

    Yamada, Tomoaki ; Kon, S.

  • Author_Institution
    Electr. & Magn. Div., AIST/NMIJ, Tsukuba, Japan
  • fYear
    2014
  • fDate
    24-29 Aug. 2014
  • Firstpage
    728
  • Lastpage
    729
  • Abstract
    This paper describes calibration methods of current comparators (CCs) and balancing current generators (BCGs) accompanied with them, for establishing a high-accuracy ac current ratio standard up to 4 kHz. The CC calibration is explained mathematically according to the self-calibration procedure that NMIJ has been exploiting. The BCG calibration is based on measurement of the conversion ratio. The technique with a CC and CT is unique in terms of the fact that the uncertainty can be ignored against the CT calibration.
  • Keywords
    calibration; current comparators; electric current measurement; mathematical analysis; measurement standards; BCG; CC; NMIJ; balancing current generator; calibration method; conversion ratio measurement uncertainty; current comparator; high-accuracy AC current ratio standard; mathematical analysis; Calibration; Current measurement; Current transformers; IEC standards; Uncertainty; Windings; Current transformer; ac current ratio standards; current comparator; self calibration;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM 2014), 2014 Conference on
  • Conference_Location
    Rio de Janeiro
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4799-5205-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2014.6898592
  • Filename
    6898592