Abstract :
The feasibility of one-dimensional phased array transducer with mechanical scanning to produce a conformal heating for external ultrasound hyperthermia was evaluated. For the configuration of this system, numbers of focal spots were distributed within the target volume. To achieve the goal of conformal heating, an algorithm was introduced to calculate the power weighting for each field pattern to produce a heating volume conformal to the shape of the target volume. Simulation results showed that rather high cure rate in target volume with less damage rate in normal tissue can be obtained for target volumes with different shapes
Keywords :
biomedical transducers; biomedical ultrasonics; gradient methods; hyperthermia; optimisation; radiation therapy; ultrasonic transducer arrays; US hyperthermia applicator; clinical hyperthermia; conformal heating; cylindrical heating volume; external ultrasound hyperthermia; field pattern; focal spot distribution; iterative gradient decent optimization; mechanical scanning; one-dimensional phased array transducer; power weighting; scanned 1D phased array; tissue; Cogeneration; Heating; Hyperthermia; Neoplasms; Phased arrays; Shape; Temperature; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2001 IEEE
Conference_Location :
Atlanta, GA
Print_ISBN :
0-7803-7177-1
Type :
conf
DOI :
10.1109/ULTSYM.2001.991968
Filename :
991968
Link To Document :
بازگشت