• DocumentCode
    236763
  • Title

    MRI heating reduction for external fixation devices using absorption material

  • Author

    Xin Huang ; Jianfeng Zheng ; Ji Chen ; Xin Wu ; Kono, Mari ; Hozono, Hideki ; Kainz, Wolfgang

  • Author_Institution
    ECE Dept., Univ. of Houston, Houston, TX, USA
  • fYear
    2014
  • fDate
    4-8 Aug. 2014
  • Firstpage
    113
  • Lastpage
    117
  • Abstract
    Using absorption materials to reduce RF heating during MRI procedure is studied in this paper. Materials with different electromagnetic absorption characteristics are used to alter the EM and current distributions on the device. Consequently, the RF induced heating behaviors of external fixation devices can be reduced. Numerical and experimental studies are provided to demonstrate the potentials of reducing the RF heating for external fixation devices of using lossy materials.
  • Keywords
    biomedical MRI; biomedical equipment; biothermics; numerical analysis; EM distributions; MRI heating reduction; MRI procedure; RF heating; absorption material; current distributions; electromagnetic absorption characteristics; external fixation devices; fixation devices of using lossy; lossy materials; numerical analysis; Absorption; Conductivity; Heating; Magnetic resonance imaging; Materials; Phantoms; Radio frequency; RF heating reduction; absorption material; external fixation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility (EMC), 2014 IEEE International Symposium on
  • Conference_Location
    Raleigh, NC
  • Print_ISBN
    978-1-4799-5544-2
  • Type

    conf

  • DOI
    10.1109/ISEMC.2014.6898953
  • Filename
    6898953