DocumentCode :
2373283
Title :
The state standard of magnetic induction unit
Author :
Zingerman, W. ; Zilbersteyn, Y. ; Bagaley, O. ; Kondratjuk, V.
Author_Institution :
State Sci. & Ind. Assoc. Metrol., Kharkov, Ukraine
fYear :
1996
fDate :
17-21 June 1996
Firstpage :
31
Lastpage :
32
Abstract :
An instrumental complex for reproduction, maintenance and transfer of the magnetic induction unit in the range from 0.05 to 2 T has been developed and researched. Homogenous magnetic fields of different values are created in a multiband electrical magnet. A special system for stabilization of the magnetic field provides standard deviation of the measurement result up to 1/spl middot/10/sup -6/. The nuclear-resonant teslameter makes it possible to measure the reproduced value with nonexcluded systematic uncertainty of 3/spl middot/10/sup -6/.
Keywords :
electromagnetic induction; magnetic fields; magnetic variables measurement; measurement errors; measurement standards; nuclear magnetic resonance; stability; units (measurement); 0.05 to 2 T; NMR; homogenous magnetic fields; magnetic induction unit; maintenance; multiband electrical magnet; nonexcluded systematic uncertainty; nuclear-resonant teslameter; reproduction; stabilization; standard deviation; state standard; transfer; Bandwidth; Coils; Control systems; Frequency; Magnetic field measurement; Magnetic fields; Magnetic resonance; Magnetic sensors; Measurement standards; Uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1996 Conference on
Conference_Location :
Braunschweig, Germany
Print_ISBN :
0-7803-3376-4
Type :
conf
DOI :
10.1109/CPEM.1996.546542
Filename :
546542
Link To Document :