Abstract :
The aim of the work is the substantiation of a calibration method of the thin-wire platinum bolometer of laser radiation pulse energy on wavelength 10.6mum. The method uses well-known table temperature dependencies of the specific resistance and the specific heat capacity of platinum. The temperature dependence of the factor of absorption efficiency is determined by temperature dependencies of the refractive and absorption indexes of bolometer material. The averaged along the bolometer length equation of heat balance with no stationary coefficients and nonuniformity distribution of incident energy is solved
Keywords :
absorption coefficients; bolometers; calibration; electric resistance; laser beams; photorefractive materials; platinum; refractive index; specific heat; 10.6 micron; Pt; absorption efficiency; bolometer material absorption indexes; bolometer material refractive index; heat balance equation; laser radiation pulse energy; specific heat capacity; specific resistance; table temperature dependencies; thin-wire platinum bolometer calibration method; Absorption; Bolometers; Calibration; Equations; Optical materials; Optical pulses; Optical refraction; Platinum; Resistance heating; Temperature dependence;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Advanced Optoelectronics and Lasers, 2003. Proceedings of CAOL 2003. First International Conference on
Conference_Location :
Alushta, Crimea
Print_ISBN :
0-7803-7948-9
Type :
conf
DOI :
10.1109/CAOL.2003.1251299
Filename :
1251299
Link To Document :
بازگشت