• DocumentCode
    23752
  • Title

    Sensing Contact Between Microneedle Array and Epidermis Using Frequency Response Measurement

  • Author

    Beelee Chua ; Peijie Cao ; Desai, Shashi P. ; Tierney, Michael J. ; Tamada, Janet A. ; Jina, Arvind N.

  • Author_Institution
    ArKal Med. Inc., Fremont, CA, USA
  • Volume
    14
  • Issue
    2
  • fYear
    2014
  • fDate
    Feb. 2014
  • Firstpage
    333
  • Lastpage
    340
  • Abstract
    This paper presents a preliminary proof of concept where contact between a microneedle array (MNA) and epidermis was sensed using frequency response measurement. The microneedles used in this paper had a height of ~ 350 μm, a pitch of 450 μm and were filled with saline. With the Randles electrode-to-electrolyte interface model, the cutoff frequency of MNA in contact with the epidermis was calculated to be significantly lower than that of the dry electrode on stratum corneum. This is in agreement with the experimental measurement of two different MNA configurations. In the first configuration, a pair of MNAs was used and in the second configuration, a MNA was paired with a wet (ECG) electrode. Using a frequency sweep from 10 Hz to 200 kHz on the first configuration, the cutoff frequency of the MNA pair was ~ 6 kHz lower than that of the dry electrode pair. Similarly in the second configuration, the MNA-ECG electrode pair and dry electrode-ECG electrode pair were differentiated by a significant peak-to-peak voltage difference of ~ 300 mV across series resistances of 10 and 100 kΩ at frequencies ~ 90-100 Hz. Through fitting the analytical and experimental frequency response plots of the MNA (R2=0.994), the component values in the Randles model were estimated to be R1=10 kΩ, C1=15 nF, and R1s=300 Ω.
  • Keywords
    biomedical electrodes; electrocardiography; frequency response; microelectrodes; needles; skin; MNA configurations; MNA-ECG electrode pair; Randles electrode-electrolyte interface model; dry electrode-ECG electrode pair; epidermis; frequency 10 Hz to 200 kHz; frequency response measurement; microneedle array; sensing contact; stratum corneum; wet ECG electrode; Electrodes; Epidermis; Equivalent circuits; Frequency measurement; Frequency response; Resistors; Microneedle array; Randles model; contact; epidermis; frequency response;
  • fLanguage
    English
  • Journal_Title
    Sensors Journal, IEEE
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2013.2283301
  • Filename
    6607215