Abstract :
A new dosimetric assessment system was conceived and implemented at Supelec. This system has performances much higher than those of the commercial systems currently available on the market. Its rate of measurement can reach 7 points per second and its sensitivity is of 15 muW/kg. Three different applications are described highlighting the interest of this system. The mobile phones dosimetry allows the comparison with commercial systems. The whole body dosimetry shows the advantage of the spped of this system, while the dosimetry of Bluetooth earpieces highlights the interest of its sensitivity.
Keywords :
Bluetooth; dosimetry; earphones; mobile radio; Bluetooth earpieces; dosimetric assessment system; mobile phones dosimetry; Application software; Bluetooth; Data processing; Dosimetry; Mobile handsets; Noise measurement; Power measurement; Probes; Time measurement; Voltage;