Title :
Performance assessment techniques for doppler radar physiological sensors
Author :
Hafner, Noah ; Lubecke, Victor M.
Author_Institution :
Electr. Eng. Dept., Univ. of Hawaii, Honolulu, HI, USA
Abstract :
This paper presents a technique for assessing the performance of continuous wave Doppler radar systems for physiological sensing. The technique includes an artificial target for testing physiological sensing radar systems with motion analogous to human heart movement and software algorithms leveraging the capabilities of this target to simply test radar system performance. The mechanical target provides simple to complex patterns of motion that are stable and repeatable. Details of radar system performance can be assessed and the effects of configuration changes that might not appear with a human target can be observed when using this mechanical target.
Keywords :
Doppler radar; biological organs; biomechanics; cardiology; motion measurement; sensors; Doppler radar physiological sensors; artificial target; complex motion patterns; continuous wave Doppler radar systems; human heart movement; physiological sensing; software algorithms; Doppler Effect; Humans; Monitoring, Physiologic; Radar;
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2009.5332661