DocumentCode :
2382176
Title :
Performance evaluation of parametric bias field correction
Author :
Zagorodnov, Vitali ; Hansen, Mads Fogtmann
Author_Institution :
Nanyang Technol. Univ., Singapore
Volume :
3
fYear :
2005
fDate :
11-14 Sept. 2005
Abstract :
Magnetic resonance (MR) images often exhibit grayscale nonuniformities, caused by radio frequency (RF) coil design or acquisition sequences. Many algorithms to remove these nonuniformities have been proposed in the past decade. However, only minor attention has been given to the performance evaluation of existing methods. We derive a link between the estimation performance and underlying image structure. For a piecewise constant 1D signal model with equal size regions we demonstrate that the variance in estimation grows as M2 , where M is the number of regions. In 2D case the growth becomes linear in M.
Keywords :
biomedical MRI; image colour analysis; MR images; RF coil design; acquisition sequences; grayscale nonuniformities; image structure; magnetic resonance images; parametric bias field correction; piecewise constant 1D signal model; radio frequency coil design; Additive noise; Biomedical imaging; Coils; Gray-scale; Image segmentation; Magnetic resonance; Magnetic resonance imaging; Partitioning algorithms; Radio frequency; Visualization;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Image Processing, 2005. ICIP 2005. IEEE International Conference on
Print_ISBN :
0-7803-9134-9
Type :
conf
DOI :
10.1109/ICIP.2005.1530496
Filename :
1530496
Link To Document :
بازگشت