Title :
Polyimide-based multi-channel arrayed electrode for measuring EEG signal on the skull of mouse
Author :
Baek, Dong-Hyun ; Lee, Eun-Joong ; Moon, Jin-Hee ; Choi, Jee Hyun ; Pak, James Jungho ; Lee, Sang-Hoon
Author_Institution :
Dept. of Biomed. Eng., Korea Univ., Seoul, South Korea
Abstract :
In this paper, we have developed 40 channel multiple electrodes mounted on the surface of mouse´s skull using polyimide substrate and tested its performance by measuring EEG signals. The recording site of the electrode was electroplated by Pt to enhance both contact impedance and adhesive strength by applying proper current, cleaning surface and removing H2 gas bubbles. For in vivo test, the electrode was placed on the skull of F1 mouse and EEG signals were measured. We observed the suitability of electrode for measuring EEG signals from multiple areas on the skull. The spectrum of EEG signal to change was observed by urethane administration.
Keywords :
adhesion; biomedical electrodes; electroencephalography; electroplating; polymers; EEG signal; adhesive strength; contact impedance; electroplating; mouse skull; multichannel arrayed electrode; polyimide; urethane; Animals; Biomedical Engineering; Electric Impedance; Electrodes; Electroencephalography; Equipment Design; Mice; Mice, Inbred C57BL; Platinum; Resins, Synthetic; Signal Processing, Computer-Assisted; Skull;
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2009.5333320