    A protocol design for evaluation of wearable cuff-less blood pressure measuring devices

  • Author

    Yan, Iris R F ; Poon, Carmen C Y ; Zhang, Y.T.

  • Author_Institution
    Dept. of Electron. Eng., Chinese Univ. of Hong Kong, Hong Kong, China
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    7045
  • Lastpage
    7047
  • Abstract
    This paper proposes one of the unique requirements in the validation protocol of the IEEE P1708 standard in assessing wearable cuff-less blood pressure (BP) measuring devices. Based on principles that are different from that of the conventional cuff-based devices, the cuff-less BP measurement approaches often require an individual calibration procedure. In this study, we used data from an experiment carried out on 28 subjects with a total of 139 sets of BP measurements as an example to show that breakdown of the performance evaluation of cuff-less devices according to the change of BP from the point of calibration is crucial for understanding and interpreting the overall accuracy of the device.
  • Keywords
    IEEE standards; biomedical electronics; blood pressure measurement; calibration; measurement standards; IEEE P1708 standard validation protocol; calibration procedure; measuement device accuracy; wearable cuff less blood pressure measuring device; Adult; Biomedical Engineering; Blood Pressure; Blood Pressure Monitoring, Ambulatory; Clinical Protocols; Equipment Design; Exercise; Humans; Young Adult;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5333380
  • Filename
    5333380