DocumentCode :
2395560
Title :
Automatic actuator control by leg load signal of active AFO for achilles tendon ruptures
Author :
Yoshizawa, Nobuyuki
Author_Institution :
Nippon Inst. of Technol., Saitama, Japan
fYear :
2009
fDate :
3-6 Sept. 2009
Firstpage :
5052
Lastpage :
5055
Abstract :
This paper describes automatic actuator control hardware for an active ankle foot orthosis (AAFO) for a ruptured Achilles tendon. The sole of the AAFO is equipped with a servomotor. The actuator can switch the upright and the forward stepping posture of the patient. To control the actuator automatically during a gate cycle, 1: joint kinematics were analyzed that consisted of both stick pictures and EMG signals from the affected limb when the patient was walking on a treadmill. The cross-correlation coefficient between the ankle angle and the EMG signal of the lower limb (tibialis anterior) was small when the patient was wearing the AAFO. 2: An air pressure sensor and a film sensor were compared experimentally to measure the leg load. A prototype AAFO with automatic control was realized by using the leg load signal provided by an air pressure sensor to realize actuator control.
Keywords :
actuators; electromyography; medical control systems; orthotics; pressure sensors; Achilles tendon ruptures; EMG signals; active AFO; active ankle foot orthosis; air pressure sensor; automatic actuator control; film sensor; joint kinematics; leg load signal; Achilles Tendon; Ankle; Biomechanics; Electromyography; Foot; Humans; Orthotic Devices; Rupture;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
Conference_Location :
Minneapolis, MN
ISSN :
1557-170X
Print_ISBN :
978-1-4244-3296-7
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2009.5333670
Filename :
5333670
Link To Document :
بازگشت