Design and performance evaluation of a high resolution small animal positron tomograph
Author :
Pavlopoulos, S. ; Tzanakos, George
Author_Institution :
Dept. of Biomed. Eng., Rutgers Univ., Piscataway, NJ, USA
fYear :
1993
fDate :
31 Oct-6 Nov 1993
Firstpage :
1697
Abstract :
The authors have used Monte Carlo simulation methods to design and evaluate a small animal positron tomograph of inner diameter 34.37 cm, field of view diameter of 20 cm, and axial field of view of 6 cm. The tomograph consists of 16 BGO blocks, each of which is partially segmented to 18×16 crystals of dimension 5.35 mm×3.30 mm×30 mm and is read out by a position sensitive photomultiplier. The rate performance as well as the spatial resolution have been studied and compared with other tomographs
Keywords :
Monte Carlo methods; biological techniques; performance evaluation; positron emission tomography; 6 to 34.37 cm; BGO blocks; Monte Carlo simulation methods; PET performance evaluation; PET system design; high resolution small animal positron tomograph; partially segmented blocks; physiological research; rate performance; spatial resolution; Animals; Cathodes; Detectors; Energy resolution; Imaging phantoms; Photomultipliers; Photonic crystals; Positron emission tomography; Spatial resolution; Timing;