• DocumentCode
    2424869
  • Title

    An 8-channel skin impedance measurement system for acupuncture research

  • Author

    Thong, Tran ; Colbert, Agatha P. ; Larsen, Adrian P.

  • Author_Institution
    Dept. of Sci. & Eng., Oregon Health & Sci. Univ., Beaverton, OR, USA
  • fYear
    2009
  • fDate
    3-6 Sept. 2009
  • Firstpage
    861
  • Lastpage
    864
  • Abstract
    An 8-channel skin impedance measurement system for acupuncture research has been developed. The underlying model of the skin used is a parallel R & C network. Pulses are used to measure the R and C values. The measurement circuit is time multiplexed across the 8 channels at the rate of 2 measurements per second, leading to a complete set of measurements every 4 seconds. In static tests, the system has been operational for over 2 days of continuous measurements. In preliminary human tests, measurements over 2 hours have been collected per subject.
  • Keywords
    RC circuits; bioelectric phenomena; biomedical measurement; electric impedance measurement; physiological models; skin; acupuncture research; eight-channel skin impedance measurement system; human tests; measurement circuit; parallel R-C network; skin model; static tests; time 2 day; time 2 h; time 4 s; Acupuncture; Amplifiers, Electronic; Electric Conductivity; Equipment Design; Equipment Failure Analysis; Humans; Plethysmography, Impedance; Reproducibility of Results; Sensitivity and Specificity; Skin Physiological Phenomena;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-3296-7
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2009.5335156
  • Filename
    5335156