Abstract :
An experimental study was carried out to evaluate the performance of a polydimethylsiloxane (PDMS) based micromixer for heterogeneous immunoassays of insulin. The detection method is based on coupling the highly specific double-antibody sandwich immunoassay with the sensitive chemiluminescence of the Luminol-Hydrogen Peroxide reaction. Both the immune and chemiluminescence reactions are conducted in the micromixer, and the resultant light emission is detected by a highly sensitive photometer. The behavior of the luminol based chemiluminescence is parabolic with respect to time, and the maximum light intensity is used here to represent the output of the reaction. The current results indicate that at an actuation pressure of 15 psi, an actuation frequency of 30 Hz leads to the best reaction result. At these operating conditions, the detection limit of insulin is about 10-14 mol/L while the total reaction time of the insulin immunoassay in the micromixer is less than 5 minutes.
Keywords :
bioMEMS; biochemistry; biosensors; chemical sensors; chemiluminescence; photometers; polymers; PDMS based micromixer; actuation pressure; chemiluminescence; double-antibody sandwich immunoassay; heterogeneous immunoassays; highly sensitive photometer; insulin; luminol-hydrogen peroxide reaction; polydimethylsiloxane; Immunoassay; Insulin; Micromixer; PDMS;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nano/Micro Engineered and Molecular Systems (NEMS), 2010 5th IEEE International Conference on
Conference_Location :
Xiamen
Print_ISBN :
978-1-4244-6543-9
Type :
conf
DOI :
10.1109/NEMS.2010.5592478
Filename :
5592478
Link To Document :
بازگشت