• DocumentCode
    2445846
  • Title

    A signal approach analysis of the Ramsey pattern in a cesium beam frequency standard

  • Author

    Makdissi, A. ; deClercq, E.

  • Author_Institution
    Lab. Primaire du Temps et des Frequences, Obs. de Paris, France
  • fYear
    1996
  • fDate
    17-21 June 1996
  • Firstpage
    271
  • Lastpage
    272
  • Abstract
    In this paper we show that the measurement of one Ramsey pattern permit us to estimate the microwave (MW) power very precisely, to calculate some of the effects that influence the accuracy of the standard without using the time of flight (TOF) distribution and to evaluate the TOF distribution very simply.
  • Keywords
    atomic beams; atomic clocks; caesium; cepstral analysis; frequency measurement; measurement standards; spectral line breadth; time of flight spectra; Cs; Rabi pulsation; Ramsey pattern; accuracy; cesium beam frequency standard; microwave power; signal approach analysis; standard; time of flight distribution; Atomic measurements; Cepstrum; Clocks; Frequency; Pattern analysis; Physics; Random variables; Resonance; Signal analysis;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1996 Conference on
  • Conference_Location
    Braunschweig, Germany
  • Print_ISBN
    0-7803-3376-4
  • Type

    conf

  • DOI
    10.1109/CPEM.1996.547067
  • Filename
    547067